Helix PCD Kit (Device and Test Strips)
Helix Process Challenge Device (PCD) for Steam Sterilisers
Includes 250 Test Strips
Helix Test Kit (Steam)
What’s Included
- 250 test strips
- High-resistance helix PCD tube
- Optional cloth bag, depending on the manufacturer
- Australia-wide shipping
- To place an order, contact our team.

Frequently Asked Questions about the Helix PCD Kit
What Does a Helix Test Do?
A Helix Process Challenge Device is used in steam sterilisers to verify effective air removal and steam penetration. This is essential for maintaining sterilisation and infection-control requirements.
Who Uses Helix Tests?
Helix tests are required for practices and clinics operating pre-vacuum, Class B autoclaves that sterilise hollow or solid loads, as outlined in AS 5369:2023.
How Often Should Helix Tests Be Performed?
Helix tests must be completed daily. Most users run them immediately after their vacuum test each morning. Refer to Australian Standard AS 5369:2023 for detailed requirements.
How Do You Perform a Helix Test?
Instructions may vary by manufacturer. The typical process is:
- Remove the cap from the device.
- Fold one test strip in half.
- Insert the strip (folded end first).
- Replace the cap.
- Place in the cloth bag if provided.
- Put the device into an empty steam steriliser.
- Run the Helix cycle on your autoclave.